Home/Services/EU MDR Medical Writing/CER

EU MDR Clinical Evaluation

Clinical Evaluation Report (CER) Writing Services for EU MDR

IMTS writes MDR-compliant Clinical Evaluation Reports to Article 61 and Annex XIV, using MEDDEV 2.7/1 Rev 4 where applicable, for Class I to Class III medical devices, including implantables, SaMD, AI-enabled devices, and legacy devices, with 300+ CERs delivered.

Reviewer ready Clinical Evaluation Report writing for medical devices, supported by rigorous literature review, evidence appraisal, and regulatory traceability.

The IMTS perspective

CER writing grounded in clinical interpretation.

Indus MedTech Solutions (IMTS) provides Clinical Evaluation Report (CER) writing services for medical device manufacturers, including Clinical Evaluation Plans (CEPs), new CERs, lifecycle updates, clinical evidence appraisal, and notified body response support. We connect the device's intended purpose and claims with the available clinical evidence, its limitations, and the conclusions the evidence can support.

A Clinical Evaluation Report brings together the device description, intended purpose, clinical context, state of the art, relevant clinical data, and benefit-risk evaluation. IMTS supports manufacturers developing a new report, updating an existing evaluation, or addressing reviewer questions. The scope may include the complete CEP and CER package or a defined task within an existing clinical evaluation program.

A well-written report cannot compensate for missing or unsuitable evidence. Our focus is to make the reasoning transparent: which data apply to the device, what they demonstrate, where results conflict, and what remains uncertain. We assess the relationship between claims, patient populations, endpoints, follow-up, and study limitations so that conclusions remain proportionate to the evidence.

Scope of support

CER services, from evaluation planning to remediation.

How we work

From source evidence to a traceable clinical evaluation.

01

Scope the device and evidence

Review the intended purpose, indications, claims, device variants, existing documentation, and reason for the project. Agree on the evidence period, deliverables, source responsibilities, review arrangements, and dependencies before drafting.

02

Evaluate and interpret the data

Develop or update the evaluation plan, review literature and manufacturer-held clinical data, and assess relevance, quality, and limitations. Distinguish device evidence from clinical background and explain the implications of unfavorable findings, conflicting results, and evidence gaps.

03

Review, reconcile, and hand over

Draft the clinical evaluation, check source traceability and consistency with related documents, and resolve comments through the agreed review rounds. Hand over the report and scoped supporting records, with outstanding inputs and future evidence needs clearly identified.

Why IMTS

Medical device expertise behind the evaluation.

Indus MedTech Solutions is a medical device writing company founded and led by Aarti Urs, PhD. Our experience includes more than 300 CERs authored and 350 systematic literature reviews conducted, supported by advanced-degree medical and regulatory writers.

We connect clinical evidence interpretation with literature review and the wider device documentation program. That means assessing whether a study supports the intended claim, making the reasoning traceable, and keeping clinical conclusions consistent across connected documents. Your project scope identifies the lead contact, review responsibilities, deliverables, and timetable.

Therapeutic experience

Experience across medical device specialties.

These therapeutic areas are examples of our experience, not an exhaustive list. IMTS also supports projects in other clinical and device areas. Contact us to discuss your device and the expertise your project requires.

Cardiovascular

Orthopedic

Dental

Spine

Oncology

Ophthalmology

Inside the evaluation

What our clinical evaluation addresses.

Device scope and clinical claims

Define the device and variants, intended purpose, indications, target population, and claimed clinical benefits. These details establish which evidence is relevant and which questions the evaluation needs to answer.

State of the art and clinical benchmarks

Evaluate the clinical condition, current treatment options, relevant comparators, and published safety and performance outcomes. Explain whether proposed benchmarks are appropriate for the device, population, endpoints, and follow-up.

Clinical data and equivalence

Identify and appraise relevant literature and manufacturer-held clinical data. When an equivalence argument is proposed, examine its supporting justification separately from evidence used to describe similar devices or the broader clinical context.

Safety, performance, and benefit-risk conclusions

Synthesize favorable and unfavorable findings, explain limitations and conflicting results, and assess what the evidence supports. Connect the clinical conclusions with the applicable safety and performance questions rather than relying on a sequence of study summaries.

Evidence gaps and lifecycle follow-up

Identify residual uncertainties and their implications for the evaluation. Connect relevant questions with PMCF objectives and postmarket findings, and keep resulting conclusions consistent with the related documentation.

Choose the right scope

New CERs, updates, and remediation need different work.

01

A new clinical evaluation

Establish the evaluation scope and plan, assess the available evidence, and develop the report with its supporting review records. Early scoping helps distinguish a writing need from an evidence-generation need before the project timetable is agreed.

02

An existing CER that needs updating

Review the previous evaluation methods, search history, current device information, and evidence generated since the last report. Assess whether changes in claims, clinical knowledge, or postmarket findings require revision of the evaluation rather than a simple extension of the literature search.

03

A CER with reviewer questions

Map each comment to the relevant evidence, analysis, or document section. Develop a focused response and revision plan, distinguishing issues that can be resolved through clarification from those requiring further data or a change in the clinical argument.

Project deliverables

Know what your CER project will include.

Evaluation plan and report

The agreed CEP and CER deliverables, with defined device scope, evidence period, evaluation methods, clinical analysis, and conclusions. The proposal states whether both documents are included and which existing materials will be reused or revised.

Literature and appraisal records

When included in scope: search strategies and dates, selection records, appraisal criteria, evidence tables, and the literature synthesis. Access to databases and full texts, reference-library format, and any supplied review records are agreed before work begins.

Traceability and evidence gaps

Supporting references and tables that connect relevant findings to the evaluation. Gaps, conflicting evidence, and unresolved inputs remain visible, with their implications explained. A separate claim-to-evidence matrix or gap assessment can be specified as a deliverable.

Review and response materials

The agreed client review rounds, comment resolution, and final handover format. The quotation also identifies any response matrix, post-submission support, or future update arrangements.

Getting started

The source documents shape the evaluation.

Start with the device description, variants, intended purpose, indications, claims, and current labeling. Existing CEPs, CERs, search strategies, appraisal tables, and reference libraries help establish what has already been evaluated and whether the previous methods remain suitable.

Clinical investigation reports and other device data need to be considered alongside relevant risk management, PMS findings, and PMCF Plans and Evaluation Reports. For updates or remediation, provide the reporting periods, device or claim changes, and reviewer questions. The scoping discussion establishes document versions, owners, outstanding inputs, and the review timetable.

The manufacturer supplies and confirms device information and controlled source records, participates in clinical and regulatory decisions, and approves the final documentation. The project scope should identify who drafts, checks, clinically reviews, and signs the evaluation; authoring support does not automatically include a separate clinical evaluator or physician review.

Evidence and quality

Explain the limits as clearly as the findings.

Quality review connects source verification with clinical reasoning. We check that cited results can be traced to their source, device descriptions and claims are consistent, evidence tables agree with the narrative, and conclusions reflect the limitations of the available data.

The project plan defines who performs the checks and how comments are resolved, including any separately commissioned independent or clinical review. Checks address the agreed evidence period and whether the conclusions reflect both favorable and unfavorable results. Differences between study populations, device versions, endpoints, or follow-up can materially change what a finding supports.

For example, a study reporting favorable outcomes at three months does not automatically support a claim about sustained performance over several years. The evaluation needs to explain the relevance of the data, identify the unsupported step in the argument, and assess what further evidence is needed. This is an illustrative appraisal question, not a client case study or an acceptance claim.

Connected evidence

Keep the CER connected to the device lifecycle.

A CER should remain consistent with its source documents and the broader evidence program. IMTS can scope connected systematic literature review, PMCF, PSUR, and SSCP writing support, with responsibilities and deliverables defined for each document.

State-of-the-art evidence for CERs

Aligning CER, PMS, PMCF, PSUR, and SSCP

Compare CER writing providers and project scopes

For regulatory context, consult the European Commission's MDCG guidance collection, including guidance on clinical evaluation equivalence and sufficient clinical evidence for legacy devices. Guidance applicability depends on the device and evaluation question.

Regulatory context

Regulatory framework we write to

IMTS develops clinical evaluations around Regulation (EU) 2017/745 Article 61 and Annex XIV Part A. These establish the legal basis for planning, appraising, analyzing, and documenting the clinical evidence supporting the device. The scope starts with its intended purpose, claims, relevant safety and performance requirements, and the level of evidence needed. A familiar report structure is useful only when it supports the required evaluation.

MEDDEV 2.7/1 Rev 4 provides established clinical evaluation methods, but it was written under the previous directives. We use it where applicable alongside the MDR and current guidance, rather than treating it as an alternative regulatory framework. MDCG 2020-5 informs equivalence assessment, MDCG 2020-6 addresses sufficient evidence for legacy devices, and MDCG 2020-13 describes the notified body clinical evaluation assessment report template. That assessment template is not a manufacturer CER template or an acceptance guarantee.

Clinical evaluation also connects with PMCF planning and reporting, addressed by MDCG 2020-7 and MDCG 2020-8. The document set should explain how postmarket findings affect the evaluation and remaining evidence questions. Applicable guidance is selected for the device and project; the MDR takes precedence where guidance and legislation differ.

MDR on EUR-Lex

MDCG guidance collection

Regulatory context

CERs by device class

IMTS supports clinical evaluation writing for Class I, IIa, IIb, and III devices, including implantables. The risk class helps frame the assessment, but it does not replace an evaluation of intended purpose, novelty, clinical claims, and available evidence. A lower classification is not a reason to omit clinical evaluation or assume that a short literature summary will establish conformity.

For Class I devices, the evaluation must still justify the evidence used. For Class IIa and IIb devices, the analysis needs to connect the clinical data with the specific indications, risks, and claimed benefits. For Class III and implantable devices, clinical investigations are generally required, subject to the conditions and exceptions in Article 61(4)–(6). An equivalence argument or legacy status requires a documented assessment; neither should be assumed to remove the need for appropriate clinical data.

Updates should follow a justified lifecycle review schedule and respond to new findings or changes. Reporting obligations are related but distinct: Class IIa PSURs are updated when necessary and at least every two years, while Class IIb and III PSURs are updated at least annually. The CER, PMCF findings, and PSUR conclusions should remain consistent without treating their update provisions as interchangeable.

Clinical investigation exceptions: MDCG 2023-7

PSUR writing support

Regulatory context

CERs for software and AI devices (SaMD)

IMTS supports clinical evaluation writing for software as a medical device, including AI-enabled devices. The evaluation needs to explain the medical purpose of the output and the evidence supporting its use in the intended population and clinical setting. A software feature list or an attractive model-performance result does not, by itself, establish that the clinical claim is supported.

MDCG 2020-1 organizes software clinical evidence around valid clinical association, technical performance, and clinical performance. The association connects the output with the targeted condition or physiological state. Technical performance concerns whether the software generates its intended output reliably from the inputs. Clinical performance addresses whether that output achieves its intended clinical purpose. These questions are connected, but the evidence for one should not automatically be presented as evidence for all three.

The CER should identify the evaluated software version, input data, intended users, clinical workflow, relevant endpoints, and evidence limitations. For AI-enabled software, we examine how the reported results relate to the claimed use and target population. Version changes, new indications, or postmarket findings may require reassessment of the evidence and the evaluation scope.

Software clinical evaluation: MDCG 2020-1

Regulatory context

Legacy device CERs and MDD-to-MDR transition

A legacy device CER update needs to reassess the evidence against MDR expectations, not simply replace references to the previous directives. IMTS reviews the existing evaluation, clinical claims, literature, investigation data, and postmarket findings to identify where the clinical argument remains supported and where further evidence or analysis is needed. Prior market history is relevant context, rather than a substitute for evaluating the quality and applicability of the data.

MDCG 2020-6 addresses sufficient clinical evidence for legacy devices. It supports a structured assessment of the available evidence and its limitations. The evaluation should explain the relevance of historical data, current clinical practice, and device-specific findings, while making unresolved gaps visible. Reliance on an old certificate or the absence of reported complaints alone does not establish an adequate MDR clinical evidence package.

For eligible devices meeting the Article 120 conditions, the extended transition ends on 31 December 2027 for Class III and specified Class IIb implantables, or 31 December 2028 for other eligible categories. These are conditional transition dates, not new application windows. The 2024 application, quality-system, and notified body agreement conditions, continued directive compliance, significant-change restrictions, and risk conditions must be considered when assessing eligibility.

Legacy clinical evidence: MDCG 2020-6

MDR transition questions and answers

Focused CER support

Choose the next step for your clinical evaluation.

The main CER service covers the complete evaluation. These focused services and working guides address specific project needs.

Direct answers

Clinical Evaluation Report writing: common questions.

How often must a CER be updated under EU MDR?

The clinical evaluation and its documentation must be updated throughout the device lifecycle using relevant PMS and PMCF data. The MDR does not prescribe one universal CER interval for every device. The evaluation plan should justify the review schedule and triggers for earlier revision. Article 61(11) specifically requires at least annual PMCF evaluation report updates for Class III and implantable devices, and annual SSCP updates if indicated; these provisions should not be confused with a blanket annual CER rule.

What is the difference between MEDDEV 2.7/1 Rev 4 and MDR requirements?

MEDDEV 2.7/1 Rev 4 is guidance developed under the previous medical device directives. Regulation (EU) 2017/745 is binding legislation. MEDDEV methods remain useful where applicable, but the CER must address MDR requirements and relevant current guidance. Where they differ, the MDR takes precedence; an MDD-era CER should not be treated as MDR-ready solely because it follows a MEDDEV structure.

Can you write a CER for a Class III or implantable device?

Yes. IMTS supports CER writing for Class III and implantable devices, including evaluation planning, literature review, appraisal of clinical evidence, updates, and notified body responses. The project must address the applicable clinical investigation requirements and any justified exceptions, alongside device-specific claims, risks, and postmarket evidence.

Do you write CERs for software as a medical device?

Yes. IMTS supports CERs for software as a medical device, including AI-enabled devices. The evaluation connects the intended medical purpose with valid clinical association, technical performance, and clinical performance, using MDCG 2020-1 where applicable. The software version, target population, clinical workflow, and limitations of the evidence are part of the evaluation scope.

What CER writing services does IMTS provide?

IMTS supports Clinical Evaluation Plans, new and updated Clinical Evaluation Reports, systematic literature reviews, state-of-the-art assessments, clinical evidence appraisal, gap analysis, and notified body response writing. A project can cover a complete evaluation package or a defined task. The proposal specifies the documents, supporting records, review rounds, responsibilities, and handover materials included.

What is the difference between a CEP and a CER?

A Clinical Evaluation Plan defines how the clinical evaluation will be conducted, including its scope, questions, evidence sources, and methods. The Clinical Evaluation Report documents the evaluation performed, the appraisal and analysis of relevant data, and the resulting conclusions. Planning and reporting should remain consistent; changes in the evaluation approach need to be reflected in the documentation.

Can IMTS update an existing CER or address notified body comments?

Yes. IMTS can assess the existing CEP, CER, supporting review records, new evidence, and reviewer feedback before defining the revision scope. Remediation may involve clearer reasoning, additional appraisal, literature updates, source reconciliation, or identifying a genuine evidence gap. A response should address the underlying question and show where supporting evidence or revised text can be found.

What documents are needed to start a CER project?

Useful inputs include the device description and variants, intended purpose and claims, instructions for use, existing CEP and CER, risk management documentation, clinical study reports, literature-review records, and relevant PMS and PMCF findings. Previous notified body comments are important for remediation work. IMTS confirms the project-specific input list and identifies missing materials during scoping.

Does CER writing include a systematic literature review?

A literature review can be included, updated, or supplied by the manufacturer for assessment. These are different scopes. The proposal should specify the review questions, sources, search period, screening and appraisal arrangements, supporting records, and final outputs. A list of references or an undocumented search is not equivalent to a reproducible review process.

Can a CER rely on evidence from similar or equivalent devices?

Similar-device literature can help establish clinical context and the state of the art. Using another device's clinical data through an equivalence argument requires a separate justification; similarity alone does not establish equivalence. IMTS can scope an assessment of the proposed evidence and supporting documentation, while identifying limitations and questions requiring additional manufacturer input or regulatory expertise.

What determines CER cost and turnaround time?

Cost and timing depend on the device scope, number of variants and indications, maturity of the existing documents, literature volume, available clinical data, evidence gaps, and review requirements. New CERs, recurring updates, and remediation projects require different levels of work. IMTS develops a project-specific quotation after reviewing the available materials and deadline, with assumptions, dependencies, and review rounds stated explicitly.

Does a completed CER guarantee notified body acceptance?

No. A writing service cannot guarantee acceptance or replace the manufacturer's responsibilities. The outcome depends on the device, underlying evidence, applicable requirements, and assessment of the full documentation. IMTS supports clear, traceable evaluation and scoped responses to questions; missing clinical evidence should be identified rather than concealed by stronger wording.

Start a conversation

Scope your CER writing or update project.

Tell us about your device, intended purpose, existing CEP and CER, available evidence, reviewer feedback, and deadline. We will help define the deliverables, source materials, review process, and practical next step.

Book a Scoping Call